A new venous infusion pathway monitoring system.

No Thumbnail Available
Date
2007-11-16
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
A new infusion catheter pathway monitoring system employing linear integrated circuits and a low power 8 bit single chip microcomputer has been developed for hospital and home use The sensor consists of coaxial three layer conductive tapes wrapped around the polyvinyl chloride infusion tube The inner tape is the main electrode which records an AC alternating current voltage induced on the patient s body by electrostatic coupling from the normal 100 volt 60 Hz AC power line wiring field in the patient s room The outside tape layer is a reference electrode to monitor the AC voltage around the main electrode The center tape layer is connected to system ground and functions as a shield The microcomputer calculates the ratio of the induced AC voltages recorded by the main and reference electrodes and if the ratio indicates a detached infusion alerts the nursing station via the nurse call system or low transmitting power mobile phone