Automated and manufacturer independent assessment of the battery status of implanted cardiac pacemakers by electrocardiogram analysis.

No Thumbnail Available
Date
2007-02-02
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
According to international standards cardiac pacemakers have to indicate the status of their batteries upon magnet application by specific stimulation patterns The purpose of this study has been to assess whether this concept can be used as a basis for automated and manufacturer independent examination of the depletion level of pacemakers in the framework of a collaborative telemedical pacemaker follow up system A prototype of such a system was developed and tested in a real clinical environment Electrocardiograms ECGs were recorded during magnet application and automatically processed to extract the specific stimulation patterns The results were used to assign each signal a corresponding pacemaker status ok replace or undefined based on the expected behavior of the devices as specified by the manufacturer The outcome of this procedure was compared to the result of an expert examination resulting in a positive predictive value of 100 for the detection of ECGs indicating pacemaker status ok The method can therefore be utilized to quickly safely and manufacturer neutrally classify cases into the categories ok and needs further checking which in a telemedical setting may be used to increase the efficiency of pacemaker follow up procedures in the future
Description
Keywords
Citation
Collections