Safe teleoperation based on flexible intraoperative planning for robot-assisted laser microsurgery.

No Thumbnail Available
Date
2013-01-31
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
This paper describes a new intraoperative planning system created to improve precision and safety in teleoperated laser microsurgeries It addresses major safety issues related to real time control of a surgical laser during teleoperated procedures which are related to the reliability and robustness of the telecommunication channels Here a safe solution is presented consisting in a new planning system architecture that maintains the flexibility and benefits of real time teleoperation and keeps the surgeon in control of all surgical actions The developed system is based on our virtual scalpel system for robot assisted laser microsurgery and allows the intuitive use of stylus to create surgical plans directly over live video of the surgical field In this case surgical plans are defined as graphic objects overlaid on the live video which can be easily modified or replaced as needed and which are transmitted to the main surgical system controller for subsequent safe execution In the process of improving safety this new planning system also resulted in improved laser aiming precision and improved capability for higher quality laser procedures both due to the new surgical plan execution module which allows very fast and precise laser aiming control Experimental results presented herein show that in addition to the safety improvements the new planning system resulted in a 48 improvement in laser aiming precision when compared to the previous virtual scalpel system
Description
Keywords
Citation
Collections